Pumps
Enteral syringe pump
Description
This syringe-driver is dedicated to the administration of enteral nutrition, only for use with syringes for enteral use. The syringe-driver is pre… Read more
Indication
Continuous enteral nutrition by syringe-driver.
Description
This syringe-driver is dedicated to the administration of enteral nutrition, only for use with syringes for enteral use.The syringe-driver is pre-calibrated with the syringes of brands C-GON (channel 2), A-VY (channel 3), VY-P (channel 4) and Nutrifit (channel 5). Additional channels can be used to calibrate other brands of syringes, by following the procedure detailed in the instructions for use (dedicated to biomedical engineers).
The syringe-driver can be used with following volume syringes: 10 ml, 20 ml, 30 / 35 ml and 50 ml.
Display range of volume delivered: 0.1 to 9 999 ml.
Occlusion detection pressure :
- High (H) : 800 mmHg +/- 200 mmHg
- Center (C) : 500 mmHg +/- 100 mmHg
- Low (L) : 300 mmHg +/- 100 mmHg
Battery voltage: DC 12 V - A fully charged battery allows to use the syringe-driver for more than 7 hours at a speed of 5 ml/hour.
The battery must be charged for minimum 4 hours.
Weight : 2.3 kg
References and Features
Channel 2 / CGON | Channel 3 / AVY | Channel 4 / VYP | Channel 5 / Nutrifit | Packaging | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Favourites | Code | 10ml | 20ml | 30/35ml | 50ml | 10ml | 20ml | 50ml | 10ml | 20ml | 50ml | 10ml | 20ml | 50ml | Units/Box | Units/Case |
SN-50C6 | 1015.102 | 1015.202 | 1015.352 | 1015.602 | 1015.103 | 1015.213 | 1015.603 | 1015.104 | 1015.204 | 1015.604 | 1015.107 | 1015.207 | 1015.607 | 1 | 1 |
Additional information
-
Contains Latex No
-
Contains animal product No
-
Pyrogen-free No
FAQ
Codes SN-50C6 / SN-50C6-A: Is this necessary to have a certified biomedical engineer to work with this syringe-driver?
The Biomedical engineer of the hospital can calibrate syringes.